A rapid, sensitive and specific high-performance liquid chromatography (HPLC) procedure for the quantification of indinavir, a potent human immunodeficiency virus (HIV) protease inhibitor, in human plasma is described. Following C 1 8 solid-phase extraction, indinavir was chromatographed on a reversed-phase C 8 column using a simple binary mobile phase of phosphate buffer-acetonitrile (60:40, v/v). UV detection at 210 nm led to an adequate sensitivity without interference from endogenous matrix components. The limit of quantification was 25 ng/ml with a 0.1 ml plasma sample. The standard curve was linear across the range from 25 to 2500 ng/ml with an average recovery of 91.4%. The mean relative standard deviations for concentrations within the standard curve ranged between 1.4 and 9.7%. Quality control standards gave satisfactory intra- and inter-assay precision (R.S.D. from 3.5 to 15.8%) and accuracy within 15% of the nominal concentration. Sample handling experiments, including HIV heat inactivation, demonstrated analyte stability under expected handling processes. The assay is suitable for the analysis of samples from adult and pediatric patients infected with HIV.