Abstract. Perforated whole-cell patch-clamp recordings obtained with nystatin are frequently used to preserve intracellular integrity. However, the perforated-patch configuration may sometimes undergo a spontaneous change into the conventional whole-cell configuration, especially when lymphocytes are investigated. The electrophysiological criteria previously described for establishing the existence of the perforated whole-cell configuration have been shown to be insufficient. Thus, the dye eosin, applied to the pipette solution, was tested as a tool for discriminating between the perforated and the conventional whole-cell configurations on rat T-lymphocytes. The dye never entered the cell from the pipette during the entire measurement in the perforated whole-cell configuration. In contrast, all cells in the conventional whole-cell configuration became red immediately after membrane rupture. Eosin barely changed the currents studied. The results suggest that eosin is a dye of choice for verifying a true perforated-patch configuration.