Mid-infrared fiber probes allow an extended use of attenuated total reflection (ATR) measurements for topical in vivo skin analysis, which were otherwise not possible with conventional sample compartment accessories. Evanescent wave spectroscopy using a flexible fiber-optic probe from silver halide fibers of square cross-section was employed for stratum corneum characterization and keratinocyte quantification on adhesive tapes. Such a method of quantifying the amount of keratin, which can be repetitively removed from the skin surface by adhesive tape application, is essential for the study of substances topically applied and penetrating into the horny layer. For calibration, the weight of keratinocytes was determined using an ultramicro-balance. Best results were obtained with difference spectroscopy and the evaluation of the amide I absorption band intensity (correlation coefficient r=0.983). Lowest amounts per cm 2 were reached for the range down to 5 μg/cm 2 . The heterogeneity in the surface density of keratinocytes clinging to the tape was investigated by microscopy, and the thickness of some individual keratinocytes was tested by ATR-microspectroscopy and atomic force microscopy.