Procedures for the electrothermal atomic absorption spectrometric determination of lead and cadmium in urine, serum and blood are developed. For serum and blood, the samples are diluted by incorporating 0.015% (w/v) Triton X-100 and 0.1% (w/v) ammonium dihydrogenphosphate to the solutions, which are then introduced directly into the furnace. A solution containing 15% (w/v) hydrogen peroxide and 0.65% (w/v) nitric acid is also introduced into the atomizer by means of a separate injection. Zeeman-based correction is recommended. Both conventional and fast-heating programs are discussed. Calibration is carried out using the standard additions method. The reliability of the procedures is checked by analyzing three certified reference materials and by recovery studies.