To examine the intratester reliability of the Spin-T goniometer, a cervical range of motion device, in a normal Indian population.Subjects comprised 30 healthy adults with mean age of 34 years (range, 18-65 years). The subjects were stabilized in the sitting position and the Spin-T goniometer mounted on the head of the subject. The study design was a within-subject repeated intratester reliability trial conducted for cervical range of motion in 6 directions of movement. Three measurements were taken in each direction (flexion, extension lateral flexion, and lateral rotation) per participant. Reliability coefficients, intraclass correlation coefficients, and 95% confidence interval were derived from repeated-measures analysis of variance (ANOVA). Where differences in ANOVA were detected, a paired t test was conducted and the typical error values and coefficient of variance were calculated.All repeated measures showed high intraclass correlation coefficients (all >0.96, P < .01). The ANOVA detected no differences between trials for all movements except rotation. The typical error values for the rotation trials did not exceed 2.5° and the coefficient of variance did not exceed 4%, which is clinically acceptable considering the normally variable cervical range of movement.In this study, the Spin-T goniometer proved to be a reliable measuring instrument for cervical range of movement in an Indian population. The use of a laser pointer fixed to the instrument ensured a consistent neutral start position