We report in this study the results of an analytical evaluation of the urinary cortisol assay using the VIDAS(R) system (Biomerieux, France). This automate performs an indirect enzyme immunoassay ending with fluorescent detection. Precision analysis results are satisfying (within- and between-run precision =< 10%). The sensitivity is 12 nmol/L and the accuracy is correct although variable. A poor correlation is observed when comparing high pressure liquid chromatography determination. This problem may depend upon the antibody specificity used by the automate. Since this automate is practical to use and performs a correct determination of cortisol on sera and urine, it may be particularly adapted in laboratories with little demand on clinicians as soon as normal ranges are established using this technique.Nous presentons ici les resultats de l'evaluation analytique du dosage du cortisol urinaire sur l'automate Vidas(R) (Biomerieux, France). Le dosage, effectue apres extraction des urines, met en jeu une reaction immunologique par competition avec revelation enzymatique et detection finale en fluorescence. La precision du systeme est satisfaisante (repetabilite et reproductibilite inferieure a 10 %) et la limite de detection est de 12 nmol/L. Les tests de dilution et de surcharge attestent d'une exactitude irreguliere mais globalement correcte. La correlation obtenue avec notre methode de dosage par chromatographie liquide haute pression en phase inverse est faible, ce probleme etant vraisemblablement lie a la specificite de l'anticorps utilise par l'automate. Le dosage du cortisol urinaire sur Vidas(R) est donc simple, rapide, de qualite analytique acceptable, et a l'avantage de s'effectuer avec les memes reactifs que ceux du cortisol serique. Il semble adapte, en premiere intention, aux laboratoires confrontes a une demande faible ou ponctuelle du clinicien, sous reserves de disposer des valuers usuelles de la cortisolurie determinee par cette technique.