The Infona portal uses cookies, i.e. strings of text saved by a browser on the user's device. The portal can access those files and use them to remember the user's data, such as their chosen settings (screen view, interface language, etc.), or their login data. By using the Infona portal the user accepts automatic saving and using this information for portal operation purposes. More information on the subject can be found in the Privacy Policy and Terms of Service. By closing this window the user confirms that they have read the information on cookie usage, and they accept the privacy policy and the way cookies are used by the portal. You can change the cookie settings in your browser.
Evaluation of the performance of the EMIT 2000 Tacrolimus assay on the Abbott Architect c8000 analyzer.Imprecision studies were performed and patient samples were assayed by EMIT assay and by LC-MS/MS.Limit of quantification was established at 2.8 μg/L. A positive bias of 17.5% between results measured on EMIT and on LC-MS/MS was detected.EMIT 2000 Tacrolimus assay is suitable for automated analyses...