A novel optical fibre reflectance sensor coupled to a multisyringe flow injection system (MSFIA) for the determination and speciation analysis of iron at trace level using chelating disks (iminodiacetic groups) is proposed. Once iron(III) has been retained onto a chelating disk, an ammonium thiocyanate stream is injected in order to form the iron(III)–thiocyanate complex which is spectrophotometrically detected at 480nm. Iron(III) is eluted with 2M hydrochloric acid so that the chelating disk is regenerated for subsequent experiments. The determination of total iron is achieved by the on-line oxidation of iron(II) to iron(III) with a suitable hydrogen peroxide stream.A mass calibration was feasible in the range from 0.001 to 0.25μg. The detection limit (3s b /S) was 0.001μg. The repeatability (RSD), calculated from nine replicates using 1ml injections of a 0.1mg/l concentration, was 2.2%. The repeatability between five chelating disks was 3.6%. The applicability of the proposed methodology in fresh and seawater samples has been proved.The proposed technique has been validated by replicate analysis (n = 4) of certified reference materials of water with satisfactory results.