The Infona portal uses cookies, i.e. strings of text saved by a browser on the user's device. The portal can access those files and use them to remember the user's data, such as their chosen settings (screen view, interface language, etc.), or their login data. By using the Infona portal the user accepts automatic saving and using this information for portal operation purposes. More information on the subject can be found in the Privacy Policy and Terms of Service. By closing this window the user confirms that they have read the information on cookie usage, and they accept the privacy policy and the way cookies are used by the portal. You can change the cookie settings in your browser.
ObjectiveTo compare left myocardial performance index (MPI) values and reproducibility using different settings and ultrasound equipment in order to standardize optimal machine settings.
MethodsLeft MPI was prospectively evaluated by one observer performing conventional Doppler in 62 fetuses (28–36 weeks of gestational age) using different settings (changing sweep speed, gain and wall motion filter (WMF)) and two different ultrasound devices (Siemens Antares, Siemens; Voluson 730 Expert, GE Medical Systems). Intraclass coefficients of agreement (ICCs) were calculated using Bland–Altman analysis...