The Infona portal uses cookies, i.e. strings of text saved by a browser on the user's device. The portal can access those files and use them to remember the user's data, such as their chosen settings (screen view, interface language, etc.), or their login data. By using the Infona portal the user accepts automatic saving and using this information for portal operation purposes. More information on the subject can be found in the Privacy Policy and Terms of Service. By closing this window the user confirms that they have read the information on cookie usage, and they accept the privacy policy and the way cookies are used by the portal. You can change the cookie settings in your browser.
We report (1) the integration of the first functioning MEMS gas chromatography system ( muGC) featuring a micropump, a micro-column, and a micro-chemiresistor sensor array; and (2) experimental demonstration of the state-of-the-art multi-vapor gas separation and detection. In particular, we report the best GC analysis data from the first micropump-driven muGC system to date: the separation and detection...
A micro gas chromatograph (muGC) capable of quantitatively analyzing the components of complex vapor mixtures at trace concentrations is described. The muGC features a micro- preconcentrator/focuser (muPCF), dual-column pressure- and temperature-programmed separation module, and an integrated array of nanoparticle-coated chemiresistors. The latest design modifications and performance data are presented...