The Infona portal uses cookies, i.e. strings of text saved by a browser on the user's device. The portal can access those files and use them to remember the user's data, such as their chosen settings (screen view, interface language, etc.), or their login data. By using the Infona portal the user accepts automatic saving and using this information for portal operation purposes. More information on the subject can be found in the Privacy Policy and Terms of Service. By closing this window the user confirms that they have read the information on cookie usage, and they accept the privacy policy and the way cookies are used by the portal. You can change the cookie settings in your browser.
PurposeTo develop and evaluate a wideband arrhythmia‐insensitive‐rapid (AIR) pulse sequence for cardiac T1 mapping without image artifacts induced by implantable‐cardioverter‐defibrillator (ICD).
MethodsWe developed a wideband AIR pulse sequence by incorporating a saturation pulse with wide frequency bandwidth (8.9 kHz) to achieve uniform T1 weighting in the heart with ICD. We tested the performance...
Both post‐contrast myocardial T1 and extracellular volume (ECV) measurements have been associated with diffuse interstitial fibrosis. The cardiovascular magnetic resonance (CMR) field is migrating towards ECV, because it is largely insensitive to confounders that affect post‐contrast myocardial T1. Despite the theoretical advantages of myocardial ECV over post‐contrast myocardial T1, systematic experimental...