The Infona portal uses cookies, i.e. strings of text saved by a browser on the user's device. The portal can access those files and use them to remember the user's data, such as their chosen settings (screen view, interface language, etc.), or their login data. By using the Infona portal the user accepts automatic saving and using this information for portal operation purposes. More information on the subject can be found in the Privacy Policy and Terms of Service. By closing this window the user confirms that they have read the information on cookie usage, and they accept the privacy policy and the way cookies are used by the portal. You can change the cookie settings in your browser.
A 45‐year‐old man with diagnosis of short QT syndrome underwent successful implantation of dual‐chamber implantable cardioverter‐defibrillator (ICD) to prevent sudden cardiac death. Ten days after implantation, the patient was treated by inappropriate ICD discharges for frequent T‐wave oversensing. After careful reprogramming of the ICD, the T‐wave oversensing was eliminated and no T‐wave oversensing or inappropriate discharge was documented during 6‐month follow‐up period. (PACE 2010; 113–116)...