The Infona portal uses cookies, i.e. strings of text saved by a browser on the user's device. The portal can access those files and use them to remember the user's data, such as their chosen settings (screen view, interface language, etc.), or their login data. By using the Infona portal the user accepts automatic saving and using this information for portal operation purposes. More information on the subject can be found in the Privacy Policy and Terms of Service. By closing this window the user confirms that they have read the information on cookie usage, and they accept the privacy policy and the way cookies are used by the portal. You can change the cookie settings in your browser.
This paper presents a new passive shimming (PS) design scheme for the correction of static magnetic-field inhomogeneities in magnetic resonance imaging (MRI) systems. In MRI, the PS procedure usually employs magnetic-field-based or harmonic-based methods to find an optimal iron-piece configuration to improve the uniformity of the static magnetic field in the imaging region. For the field-based PS...
This paper presents a finite difference method for the design of gradient coil in MRI. In this method, a linear matrix equation is formulated using a finite-difference approximation of the current density in the source domain and an optimization procedure is then carried out to solve the resulting inverse problem and the coil winding pattern are found. The developed algorithm is tested with a transverse...