The Infona portal uses cookies, i.e. strings of text saved by a browser on the user's device. The portal can access those files and use them to remember the user's data, such as their chosen settings (screen view, interface language, etc.), or their login data. By using the Infona portal the user accepts automatic saving and using this information for portal operation purposes. More information on the subject can be found in the Privacy Policy and Terms of Service. By closing this window the user confirms that they have read the information on cookie usage, and they accept the privacy policy and the way cookies are used by the portal. You can change the cookie settings in your browser.
OBJECTIVESA prospective study to evaluate agreement and precision of a new point‐of‐care portable analyser, the EPOC analyser, compared with the i‐STAT analyser in canine blood.
METHODSBlood samples (68 venous and 32 arterial) were obtained from 63 client‐owned dogs for clinical reasons and surplus blood was used to analyse agreement between the EPOC and i‐STAT analysers. Precision of the EPOC analyser was also assessed by repeat analysis of 20 samples. Measured analytes included pH, partial pressures of carbon dioxide and oxygen and concentrations of sodium, potassium, ionised calcium, glucose and haematocrit. Haemoglobin, base excess, bicarbonate and saturation of haemoglobin with oxygen were calculated...