PurposeThe purpose of this study is to present a strategy to calculate the implant‐friendly (IF) excitation modes—which mitigate the RF heating at the contacts of deep brain stimulation (DBS) electrodes—of multichannel RF coils at 7 T.
MethodsAn induced RF current on an implantable electrode generates a scattered magnetic field whose left‐handed circularly polarizing component (
...