The Infona portal uses cookies, i.e. strings of text saved by a browser on the user's device. The portal can access those files and use them to remember the user's data, such as their chosen settings (screen view, interface language, etc.), or their login data. By using the Infona portal the user accepts automatic saving and using this information for portal operation purposes. More information on the subject can be found in the Privacy Policy and Terms of Service. By closing this window the user confirms that they have read the information on cookie usage, and they accept the privacy policy and the way cookies are used by the portal. You can change the cookie settings in your browser.
Purpose
A multi‐coil shim setup is designed and optimized for human brain shimming. Here, the size and position of a set of square coils are optimized to improve the shim performance without increasing the number of local coils. Utilizing such a setup is especially beneficial at ultrahigh fields where B0 inhomogeneity in the human brain is more severe.
Methods
The optimization started with a symmetric...
Purpose
A 16‐channel multi‐coil shimming setup was developed to mitigate severe B0 field perturbations at ultrahigh field and improve data quality for human brain imaging and spectroscopy.
Methods
The shimming setup consisted of 16 circular B0 coils that were positioned symmetrically on a cylinder with a diameter of 370 mm. The latter was large enough to house a shielded 18/32‐channel RF transceiver...